144 Stratos LV/LV-T Technical Manual
6. Pass non-absorbable ligature through
the hole in the connector receptacle to
secure the CRT-P in the pocket.
CAUTION
Tightening Setscrew(s) – Do not overtighten the
setscrew(s). Use only the BIOTRONIK supplied torque
wrench.
Sealing System – Be sure to properly insert the torque
wrench into the perforation at an angle perpendicular to the
connector receptacle. Failure to do so may result in damage
to the plug and its self-sealing properties.
NOTE:
Do not lubricate the grommets.